MCA527
Name CMD_QUERY_UF6_ROIS
Compatibility Identical to the MCA166 command.
Execution right Not necessary
Description The command reads begin and end of up to three ROIs that can be set by the command
CMD_SET_UF6_ROIS. The name of the command refers to its original usage. See also the
command CMD_QUERY_UF6_INFO.
Format integer integer integer long integer
Parameter Preamble command 0 0 end flag
Byte string (HEX) A5 5A 65 00 00 00 00 00 00 00 B9 9B
Result data array
Byte offset 0 1
st
ROI begin unsigned long
Byte offset 4 1
st
ROI end unsigned long
Byte offset 8 2
nd
ROI begin unsigned long
Byte offset 12 2
nd
ROI end unsigned long
Byte offset 16 3
rd
ROI begin unsigned long
Byte offset 20 3
rd
ROI end unsigned long
Byte offset 24 unused 82 bytes
Byte offset 106 Command flag and parameters 8 bytes
Byte offset 114 unused 12 bytes
Byte offset 126 Checksum unsigned short
Byte offset 128 unused 4 bytes